Electronic portal imaging devices (EPIDs) are potentially valuable tools for linear accelerator quality assurance and for measuring and analyzing geometric variations in radiation treatment delivery. Geometric analysis is more robust if referenced against an absolute position such as the isocenter (collimator axis of rotation), allowing the observer to discriminate between various setup errors and jaw or multileaf collimator (MLC) calibration errors. Unfortunately, mechanical instabilities in EPIDs make such analysis difficult. In the present work, we describe how MLC interleaf radiation leakage, hidden in the background of portal images, can be extracted and analyzed to find the field isocenter perpendicular to leaf travel direction. The signal from the interleaf radiation leakage is extracted to provide a precise and accurate determination of the isocenter location in the direction perpendicular to MLC leaf travel. In the direction of leaf travel, the minimization of residuals between planned and measured leaf positions is used to determine the isocenter. This method assumes that leaf positioning errors are randomly distributed. The validity of the method for determining the angular deviation between EPID image grid lines and collimator angle and for determining the known isocenter position is experimentally established.PACs numbers: 87.53.Oq, 87.53.Xd, 87.57.NK
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.