Inter-centre amplitude variability was typically within clinically significant thresholds, suggesting that inter-centre variability with suitable standardisation may not add more to total variability than inter-subject variability. Variability improvements gained by the tighter specifications of the ISCEV-specified protocol were possibly more than lost due to imprecisions of xenon flashtubes. Peak time variability was far lower than amplitude variability, corresponding with acceptable variability of biochemical assays. These results represent a vindication of the existence of an ERG standard and suggest that further standardisation would lend itself to greater reproducibility of ERGs worldwide.