A new type of subtraction imaging system was developed using an iodine filter and a single-energy broad bandwidth monochromatized x ray. The x-ray images of coronary arteries made after intravenous injection of a contrast agent are enhanced by an energy-subtraction technique. Filter chopping of the x-ray beam switches energies rapidly, so that a nearly simultaneous pair of filtered and nonfiltered images can be made. By using a high-speed video camera, a pair of two 512 × 512 pixel images can be obtained within 9 ms. Three hundred eighty-four images (raw data) are stored in a 144-Mbyte frame memory. After phantom studies, in vivo subtracted images of coronary arteries in dogs were obtained at a rate of 15 images/s.