Large detection area and good timing resolution of Silicon Photomultiplier (SiPM) based scintillation detectors is required for TOF-PET imaging. Achieving good timing requires output pulse uniformity on each microcell to be good; this depends on minimizing the gain, the recovery time, the trace impedance variation, etc. A double metal trace, a metal resistor and a new structure were applied to produce uniform characteristics on a single Multi-Pixel Photon Counter (MPPC) channel. The resulting time resolution has been improved to 140 ps for approximately 10 photons irradiation and an overvoltage of 2.5 V on a 6x4 mnl single channel MPPC test pattern. Furthermore aj'terpulse probability has been dramatically suppressed to a typical value of 3% by applying new silicon wafers and new process conditions.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.