A liquid chromatography-mass spectrometry (LC-MS/MS) method using hydrophilic interaction liquid chromatography (HILIC) was developed for the analysis of neurotoxins β-N-methylamino-L-alanine (BMAA) and 2,4-diaminobutyric acid (DAB), using multiple reaction monitoring (MRM) scan mode. Oasis-MCX and Strata-X-C polymeric cation-exchange cartridges were used to clean extracts of cyanobacterial cultures, including two strains of Microcystis aeruginosa and one strain of Nostoc sp. The performance of the solid-phase extraction (SPE) cartridges for BMAA and DAB were evaluated using mixed standards and spiked cyanobacterial extracts, which demonstrated recoveries of BMAA and DAB ranging from 66% to 91%. Matrix effects in LC-MS/MS were evaluated, and while there was no effect on BMAA quantitation, suppression of DAB was found. Full scan (Q1) and enhanced product ion (EPI) monitoring showed that the DAB suppression may be due to closely eluting compounds, including lysine, histidine, arginine and three other compounds with [M + H](+) m/z of 88, 164 and 191. The procedures developed allow the sensitive and effective analysis of trace BMAA and DAB levels in cyanobacteria. While DAB was confirmed to be present, no BMAA was found in the cyanobacterial samples tested in the present study.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.