In model experiments the precision of measurement of laser ablation depth by one A scan was 5-20 microm, depending on the signal-to-noise ratio (SNR), whereas the precision of measurement of laser ablation depth as the integral characteristic of the ablated layer profile was 0.3-5 microm. The experimental results showed that at small SNR Fourier filtering might considerably increase reliability and precision of measurements. When SNR is high, the measurement precision does not change. The precision of measurements of the corneal thickness in preliminary in vivo experiments was higher than in ex vivo experiments. This factor is very promising for application of the method suggested herein in refractive surgery.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.