The purpose of this systematic review was to identify and summarize clinical studies related to the fracture of zirconia abutments in implant treatments.Material and Methods: Medline, Embase, and Cochrane library searches were performed and complemented by manual searches from database inception to February 11, 2018, for title and abstract analysis.Results: Initially, 645 articles were obtained through database searches. Fifty-three articles were selected for full-text analysis, and 15 studies met the inclusion criteria. The selected studies were analyzed regarding fracture rate, abutment-implant connection, time point of fracture, location of critical crack, causes, managements, and preventive measures with respect to zirconia abutment fracture.Conclusions: Lower fracture rates were reported for internal connection with metal component (2-piece) zirconia abutments compared with external and internal full-zirconia connection (onepiece) zirconia abutments. Overpreparation and overload should be avoided in case of zirconia abutments.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.