In the scanning transmission x-ray microscope (STXM) the specimen is scanned in a raster by an x-ray probe formed with a Fresnel zone plate. To achieve near diffraction limited resolution it is necessary to have a coherent source, even when forming an incoherent brightfield image by measuring the x-ray intensity transmitted by the object. This has so far been the only imaging mode used in the STXM and for hydrated biological specimens is well suited to soft x-ray wavelengths within the “water window” (2·33 to 4·36 nm) where carbon absorbs much more strongly than water. However, by the use of phase contrast rather than amplitude contrast, it is possible to form images at wavelengths where the absorption is low, resulting in lower radiation dose for the same level of contrast. Calculations made by Howells [1] and Rudolph and Schmahl [2] have demonstrated very clearly the advantages of phase contrast imaging at wavelengths outside the water window.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.