Re-usable air/water and suction valves used in endoscopes often demonstrate risk of infection. To the authors' knowledge, the safety and efficacy of re-usable and single-use valves have not been compared to date. As such, a laboratory investigation was undertaken to compare the safety and efficacy of re-usable and single-use valves at 11 Italian endoscopy sites. Safety was evaluated by analysing the rinse liquid of reprocessed re-usable valves ready for use, and efficacy was assessed based on the completion of endoscopic procedures without valve malfunction. This study found significantly lower contamination of single-use valves compared with re-usable valves (0 vs 29.1%, respectively;
P
=0.007) and similar efficacy (97.6 vs 98.8%, respectively;
P
=ns). Microbiological analysis of the rinse liquid of reprocessed re-usable valves identified various surviving micro-organisms and highlighted their potential pathogenicity. Such data suggest that sterile single-use valves may be safer than re-usable valves, and have comparable performance.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.