A method for the determination of bisphenol A according to the European Union guideline, which establishes a limit of 0.1 ng/mL for organic pollutants in water, is proposed. The method involves a micro liquid-liquid extraction using dichloromethane followed by a silylation step. Identification and quantitation are performed with gas chromatography-mass spectrometry, using an HP-5MS column. The retention time is 7.02 min. Quantitation is carried out using single-ion monitoring (SIM) at m/z 73, 357, and 372. A clean-up is not necessary using SIM mode. Deuterated anthracene (2H10-anthracene) is used as an internal standard. The method is applied to the determination of bisphenol A at very low concentration levels (10.0-250.0 ng/L) in different types of natural water samples. The detection limit obtained is 0.4 ng/L. Recovery efficiencies are close to 100% in all cases.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.