Stir-bar sorptive extraction in combination with an in situ derivatisation reaction and thermal desorption-gas chromatography-mass spectrometry was successfully applied to determine parabens (methylparaben, isopropylparaben, n-propylparaben, butylparaben and benzylparaben), triclosan and methyltriclosan in water samples. This approach improves both the extraction efficiency and the sensitivity in the GC in a simple way since the derivatisation reaction occurs at the same time as the extraction procedure. The in situ derivatisation reaction was carried out with acetic anhydride under alkaline conditions. Thermal desorption parameters (cryofocusing temperature, desorption flow, desorption time, desorption temperature) were optimised using a Box-Behnken experimental design. All the analytes gave recoveries higher than 79%, except methylparaben (22%). The method afforded detection limits between 0.64 and 4.12 ng/L, with good reproducibility and accuracy values. The feasibility of the method for the determination of analytes in water samples was checked in tap water and untreated and treated wastewater.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.