An international comparison of two I 2 -stabilized frequency-doubled Nd:YAG lasers at λ 532 nm has been performed between the Istituto di Metrologia "G. Colonnetti" (IMGC)/Politecnico di Milano (PoliMI) and the Physikalisch-Technische Bundesanstalt (PTB). Both lasers used saturated absorption of the P(54) 32-0 iodine transition as a frequency reference. The signals for stabilizing the lasers were obtained using the frequency modulation spectroscopy technique with an external iodine cell. A minimum value for the relative Allan standard deviation of 7 10 -14 was reached after an integration time of 40 s. The hyperfine line separations were measured by the beat-frequency method and compared with the frequency values recently reported by Ye and co-workers. Three different iodine cells were investigated. The range of frequency differences for the iodine cells IMGC QU and PTB 1/98 between the two lasers was from -9 kHz to 7.3 kHz with a mean of 1.2 kHz and a standard deviation of 3.5 kHz. Taking into account the measured frequency offsets of the different iodine cells, the evaluated frequency repeatability of the optical standards was within 2.5 kHz.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.