This study evaluates the influence of 2 commercially available dental implant systems on stress distribution in the prosthesis, abutment, implant, and supporting alveolar bone under simulated occlusal forces, employing a finite element analysis. The implants and abutments evaluated consisted of a stepped cylinder implant connected to a screw-retained, internal, hexagonal abutment (system 1) and a conical implant connected to a solid, internal, conical abutment (system 2). A porcelain-covered, silver-palladium alloy was used as a crown. In each case, a simulated, 100-N vertical load was applied to the buccal cusp. A finite element model was created based on the physical properties of each component, and the values of the von Mises stresses generated in the prosthesis, abutment, implant, and supporting alveolar bone were calculated. In the prostheses, the maximum von Mises stresses were concentrated at the points of load application in both systems, and they were greater in system 1 (148 N/mm2) than in system 2 (55 N/mm2). Stress was greater on the abutment of system 2 than of system 1 on both the buccal (342 N/mm2 x 294 N/mm2) and lingual (294 N/mm2 x 148 N/ mm2) faces. Stress in the cortical, alveolar bone crest was greater in system 1 than in system 2 (buccal: 99.5 N/mm2 x 55 N/mm2, lingual: 55 N/mm2 x 24.5 N/mm2, respectively). Within the limits of this investigation, the stepped cylinder implant connected to a screw-retained, internal hexagonal abutment produces greater stresses on the alveolar bone and prosthesis and lower stresses on the abutment complex. In contrast, the conical implant connected to a solid, internal, conical abutment furnishes lower stresses on the alveolar bone and prosthesis and greater stresses on the abutment.