We developed an automated analytic system for glycated lipoprotein using high-performance liquid chromatography (HPLC) with an affinity boronate column and a gel permeation column. This system can measure glycated lipoprotein (glycated LDL and glycated HDL) in a small (5 microliters) sample of serum in a short time (40 min/sample). Recovery with this system was 92.1%. Therefore a large number of samples can be measured in clinical use. The system should contribute to an elucidation of the role of glycated lipoprotein in atherosclerosis.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.