Reticulated vitreous carbon (RVC), and Nafion membrane are used to fabricate a composition electrode to measure nitric oxide (NO) concentration amperometrically in the gas phase. Limit of detection was found to be 6 ppb at an applied voltage of 0.66 V (vs. mercury sulfate reference electrode) with average response time of less than 30 seconds. The response of the sensor was linearly dependent on the concentration over the whole tested range from 19 ppb-50 ppm of NO. Simplicity in electrode fabrication and consistent performance between individual sensors make RVC and Nafion attractive materials for detecting very low levels of nitric oxide gas in routine analysis.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.