It allowed quantitative measurement of CK activity from 0.5 to 250mU/ml. The accuracy of measurement was relatively high, with CV of 3% to 5%. As for interference of coexisting components in the serum, neither creatinine nor bilirubin had any effect on the CK activity levels, although the presence of 250mg/dl or more of hemoglobin in serum caused errors in the measurement of CK activity. The present thin layer assay system was found to be a feasible new method for bio-and chemi-luminescent reactions in routine laboratory use.