This paper presents an impedance plethysmography system suitable to perform a continuous monitoring of human breath activity. The problem of motion artifact is mitigated through the use of a correction technique exploiting an additional inertial sensor able to detect movements of the arms of the subject under test. The correction algorithm is based on a simple correlation technique and only requires a very brief training at the beginning of the acquisition session, with the monitored subject performing random movements in apnea condition. Application of the proposed system to a healthy adult volunteer demonstrates the potentiality of the correction algorithm, which, thanks to its extreme simplicity and low computational cost, is a suitable candidate for implementation in a low-cost and portable monitoring system
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.