The process of modern
cardiovascular device fabrication should
always be associated with an investigation of how surface properties
modulate its hemocompatibility through plasma protein adsorption as
well as blood morphotic element activation and adhesion. In this work,
a package of novel assays was used to correlate the physicochemical
properties of thin ceramic coatings with hemocompatibility under dynamic
conditions. Different variants of carbon-based films were prepared
on polymer substrates using the magnetron sputtering method. The microstructural,
mechanical, and surface physicochemical tests were performed to characterize
the coatings, followed by investigation of whole human blood quality
changes under blood flow conditions using the “Impact R”
test, tubes’ tester, and radial flow chamber assay. The applied
methodology allowed us to determine that aggregate formation on hydrophobic
and hydrophilic carbon-based coatings may follow one of the two different
mechanisms dependent on the type and conformational changes of adsorbed
blood plasma proteins.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.