Background
A more sensitive transthoracic color Doppler technology (convergent color Doppler), along with a heart rate (HR) reduction and new tomographic planes, can greatly improve coronary blood flow velocity (BFV) recordings in the left main (LMCA) and left anterior descending (LAD) coronary arteries, allowing the detection of even a slight acceleration of BFV due to mild coronary stenosis.
Methods
A group of 26 patients underwent convergent color Doppler transthoracic echocardiography (CC-Doppler TTE) in the LMCA and in the LAD coronary arteries before and after HR lowering. A second group of 71 patients scheduled for intravascular ultrasound, expanded to 96 with 25 more patients with normal LAD (by angiography/low likelihood of disease), underwent BFV Doppler recordings by CC-Doppler TTE of the whole LAD (specifically the proximal, mid and distal segments) to detect a localized increase in BFV, after attaining maximal and reference BFV in each segment.
Results
In the first group, HR reduction dramatically improved the detection of optimal flow in the LMCA and LAD, from 4 to 54% and from 6 to 94% of the segments, respectively (P < 0.001). In the second group intravascular ultrasonography (IVUS) showed mild stenoses in 60 patients. The maximum velocity was higher in the diseased segment than normal segments (49 ± 24 vs. 33 ± 11 cm/s; P < 0.001) and as the reference velocity was similar (32 ± 9 vs. 33 ± 11 cm/s; P = ns), the % increase was also higher (52 ± 52 vs. 0.7 ± 3%; P < 0.001). Using a >21% increase in velocity as a cutoff value, the sensitivity and specificity of CC-Doppler TTE in detecting at least one LAD plaque were 87% (52/60 patients [pts]) and 100% (36/36 pts), respectively.
Conclusion
CC-Doppler TTE evaluation of LAD BFV is greatly improved after reducing HR, allowing accurate noninvasive assessment of mild LAD stenosis with no radiation exposure.
Blood flow velocity evaluation in the LAD by means of transthoracic ED is feasible and reliable in detecting, locating, and assessing the severity of LAD stenosis.
OPG levels are higher in chronic CAD patients. CCA-IMT confirmed its importance in predicting CAD, showing significantly higher values in the patients in the ACS group as compared with those in the chronic CAD group.
Background: A lead-reactive fibrous capsule (FC) identified by ultrasounds as an atrial or ventricular lead thickness of more than 1 mm above the vendor-declared lead diameter (TL) and its local fibrotic attachment to the cardiac wall (FAC) have never been investigated in vivo, so their relationship with post-extraction masses (ghost) is not known. Methods: Intracardiac echocardiography (ICE) was performed twice during the same extraction procedure in 40 consecutive patients: before and immediately after infected lead extraction Results: The ghost detection rate was high: 60% (24/40 patients); ICE could identify both TL and FAC, TL being noted in 25/40 (62%) patients and FAC in 12/40 patients (30%). Both TL and FAC were significantly associated with ghosts (p < 0.001 and p = 0.002, respectively), but TL had a higher prediction power. The specificity was similar: 94% (15/16) and 100% (16/16), respectively, but TL showed a much higher sensitivity: 100%, (24/24) vs 50% (12/24) (p = 0.016). The ghost group did not show a higher event rate in the follow-up (mean follow-up time = 20 ± 17 months). Conclusion: ICE is able to evaluate both TL and FAC in vivo; ghosts are mostly benign remnants of fibrotic lead capsule cut off during extraction and retained inside the heart by FAC.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.