To assess spatial and temporal pressure characteristics in patients with repaired aortic coarctation compared to young healthy volunteers using time-resolved velocity-encoded three-dimensional phase-contrast magnetic resonance imaging (4D flow MRI) and derived 4D pressure difference maps. After in vitro validation against invasive catheterization as gold standard, 4D flow MRI of the thoracic aorta was performed at 1.5T in 13 consecutive patients after aortic coarctation repair without recoarctation and 13 healthy volunteers. Using in-house developed processing software, 4D pressure difference maps were computed based on the Navier-Stokes equation. Pressure difference amplitudes, maximum slope of pressure amplitudes and spatial pressure range at mid systole were retrospectively measured by three readers, and twice by one reader to assess inter- and intraobserver agreement. In vitro, pressure differences derived from 4D flow MRI showed excellent agreement to invasive catheter measurements. In vivo, pressure difference amplitudes, maximum slope of pressure difference amplitudes and spatial pressure range at mid systole were significantly increased in patients compared to volunteers in the aortic arch, the proximal descending and the distal descending thoracic aorta (p < 0.05). Greatest differences occurred in the proximal descending aorta with values of the three parameters for patients versus volunteers being 19.7 ± 7.5 versus 10.0 ± 2.0 (p < 0.001), 10.9 ± 10.4 versus 1.9 ± 0.4 (p = 0.002), and 8.7 ± 6.3 versus 1.6 ± 0.9 (p < 0.001). Inter- and intraobserver agreements were excellent (p < 0.001). Noninvasive 4D pressure difference mapping derived from 4D flow MRI enables detection of altered intraluminal aortic pressures and showed significant spatial and temporal changes in patients with repaired aortic coarctation.
Three-dimensional velocity-encoded cine magnetic resonance imaging (3D VEC MRI) allows for calculation of secondary flow parameters that may be used to estimate prognosis of individual cardiovascular diseases. However, its accuracy has not been fully investigated yet. The purpose of this study was to validate aortic flow quantification by 3D VEC MRI in vitro and in vivo using stacked two-dimensional acquisitions. Time-resolved stacks of two-dimensional planes with three-directional velocity-encoding (stacked-2D-3dir-MRI) were acquired in an elastic tube phantom with pulsatile flow simulating aortic flow as well as in 11 healthy volunteers (23 ± 2 years). Previously validated two-dimensional through-plane VEC MRI at six equidistant levels in vitro and three locations in vivo (ascending aorta/aortic arch/descending aorta) was used as reference standard. The percentage difference of the stacked-2D-3dir-MRI measurement to the reference standard was defined as the parameter for accuracy. For in vitro aortic flow, stacked-2D-3dir-MRI underestimated average velocity by -6.8% (p < 0.001), overestimated average area by 13.6% (p < 0.001), and underestimated average flow by -7.4% (p < 0.001). Accuracy was significantly higher in the field of view centre compared to off-centre (p = 0.001). In vivo, stacked-2D-3dir-MRI underestimated average velocity (all three locations p < 0.001) and overestimated average area at all three locations (p = n.s./<0.001/<0.001). Average flow was significantly underestimated in the ascending aorta (p = 0.035), but tended to be overestimated in the aortic arch and descending aorta. In conclusion, stacked-2D-3dir-MRI tends to overestimate average aortic area and to underestimate average aortic velocity, resulting in significant underestimation of average flow in the ascending aorta.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.