Purpose: Studies have evaluated the viability of using open-face masks as an immobilization technique to treat intracranial and head and neck cancers. This method offers less stress to the patient with comparable accuracy to closed-face masks. Open-face masks permit implementation of surface guided radiation therapy (SGRT) to assist in positioning and motion management. Research suggests that changes in patient facial expressions may influence the SGRT system to generate false positional corrections. This study aims to quantify these errors produced by the SGRT system due to face motion. Methods: Ten human subjects were immobilized using open-face masks. Four discrete SGRT regions of interest (ROIs) were analyzed based on anatomical features to simulate different mask openings. The largest ROI was lateral to the cheeks, superior to the eyebrows, and inferior to the mouth. The smallest ROI included only the eyes and bridge of the nose. Subjects were asked to open and close their eyes and simulate fear and annoyance and peak isocenter shifts were recorded. This was performed in both standard and SRS specific resolutions with the C-RAD Catalyst HD system. Results: All four ROIs analyzed in SRS and Standard resolutions demonstrated an average deviation of 0.3 AE 0.3 mm for eyes closed and 0.4 AE 0.4 mm shift for eyes open, and 0.3 AE 0.3 mm for eyes closed and 0.8 AE 0.9 mm shift for eyes open. The average deviation observed due to changing facial expressions was 1.4 AE 0.9 mm for SRS specific and 1.6 AE 1.6 mm for standard resolution. Conclusion: The SGRT system can generate false positional corrections for face motion and this is amplified at lower resolutions and smaller ROIs. These errors should be considered in the overall tolerances and treatment plan when using openface masks with SGRT and may warrant additional radiographic imaging.
Introduction
Two end‐to‐end tests evaluate the accuracy of a surface‐guided radiation therapy (SGRT) system (CRAD Catalyst HD) for position verification in comparison to a stereoscopic x‐ray imaging system (Brainlab Exactrac ) for single‐isocenter, multiple metastases stereotactic radiosurgery (SRS) using 3D polymer gel inserts.
Materials and methods
A 3D‐printed phantom (Prime phantom, RTsafe PC, Athens, Greece) with two separate cylindrical polymer gel inserts were immobilized in open‐face masks and treated with a single isocentric, multitarget SRS plan. Planning was done in Brainlab (Elements) to treat five metastatic lesions in one fraction, and initial setup was done using cone beam computed tomography. Positional verification was done using orthogonal X‐ray imaging (Brainlab Exactrac) and/or a surface imaging system (CRAD Catalyst HD, Uppsala, Sweden), and shift discrepancies were recorded for each couch angle. Forty‐two hours after irradiation, the gel phantom was scanned in a 1.5 Tesla MRI, and images were fused with the patient computed tomography data/structure set for further analysis of spatial dose distribution.
Results
Discrepancies between the CRAD Catalyst HD system and Brainlab Exactrac were <1 mm in the translational direction and <0.5° in the angular direction at noncoplanar couch angles. Dose parameters (DMean%, D95%) and 3D gamma index passing rates were evaluated for both setup modalities for each planned target volume (PTV) at a variety of thresholds: 3%/2 mm (Exactrac≥93.1% and CRAD ≥87.2%), 5%/2 mm (Exactrac≥95.6% and CRAD ≥94.6%), and 5%/1 mm (Exactrac≥81.8% and CRAD ≥83.7%).
Conclusion
Dose metrics for a setup with surface imaging was found to be consistent with setup using x‐ray imaging, demonstrating high accuracy and reproducibility for treatment delivery. Results indicate the feasibility of using surface imaging for position verification at noncoplanar couch angles for single‐isocenter, multiple‐target SRS using end‐to‐end quality assurance (QA) testing with 3D polymer gel dosimetry.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.