A coin-shaped ionisation chamber, orientated with its thin window facing away from the radiation source, was used to investigate the dose perturbations caused by the absence of back-scattering material near the exit surface of solid phantoms. Cobalt 60 and 4, 8 and 16 MV X-ray beams were used in the study. With no scattering material beyond the chamber window the ionisation was found to be as much as 17% less than the full scatter value. This was attributed to the absence of both back-scattered electrons and back-scattered photons. Full electron back-scattering could be restored by placing between 1.0 and 2.7 mm of unit density material beyond the chamber, depending on the primary beam energy. Under these circumstances the reduction in dose, now due to the absence of back-scattered photons only, was found to be small.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.