We developed a new quantitative system for diagnosis of invasive pulmonary aspergillosis (IPA) using real-time automated polymerase chain reaction (PCR). Intra-assay and interassay precision rates for in vitro examination were 2.53% and 2.20%, respectively, and the linearity of this assay was obtained when there were >20 copies/well. We examined 323 samples taken from 122 patients with hematological malignancies, including 33 patients with IPA and 89 control patients. Blood samples were subjected to PCR antigen detection methods, using enzyme-linked immunosorbent assay (ELISA) and determination of plasma (1-->3)-beta-D-glucan (BDG) concentration. The sensitivities of PCR, ELISA, and BDG measurement for diagnosis of IPA were 79%, 58%, and 67%, respectively; the specificities were 92%, 97%, and 84%. Positive findings on PCR preceded those of computed tomography by -0.3+/-6.6 days, those of BDG measurement by 6.5+/-4.9 days, and those of ELISA by 2.8+/-4.1 days. Real-time PCR was sensitive for IPA diagnosis, and quantitation was accurate.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.